Feature | April 06, 2012

St. Jude Medical Discontinues Two CRT Leads, Identifies Issue With External Conductors

April 6, 2012 — St. Jude Medical announced it is proactively informing physicians about visual observations of externalized conductors on the silicone end of QuickSite and QuickFlex left ventricular (LV) leads, used to connect cardiac resynchronization therapy (CRT) devices to the heart.

There have been no reports of patient injury or loss of therapy due to externalized conductors in these leads, but as a conservative measure, St. Jude Medical is communicating with physicians about the incidence rate so they have the most updated lead performance information with which to make important patient care decisions. The company will no longer sell the lead models and notes that the overall safety and reliability of QuickSite and QuickFlex leads continues to be comparable to currently available CRT leads from other manufacturers.

A copy of the physician letter and additional information about the leads can be found at sjmprofessional.com.

St. Jude Medical has confirmed 39 cases of externalized conductors, out of 171,000 QuickSite and QuickFlex leads sold worldwide, resulting in a current reported incidence rate of 0.023 percent, or 2.3 in 10,000. Because these leads continue to function normally, the company expects that this rate is under-reported. Based on an analysis of leads returned to the company and recent fluoroscopic images of implanted leads still in clinical use, St. Jude Medical estimates that 3 to 4 percent of QuickSite and QuickFlex leads may exhibit externalized conductors. As a result of this estimated rate, the company felt it was prudent to communicate with physicians about the externalized conductors at this time.

This medical device advisory does not affect the company’s continued sale of its newer QuickFlex µ (micro) or Quartet LV leads, which are fully insulated using Optim insulation along the entire length of the lead body. Optim insulation is a hybrid material that combines the biostability and flexibility of high-performance silicone rubber with the strength, tear resistance and abrasion resistance of polyurethane to provide increased durability and flexibility. There have been no reports or observations of externalized conductors in these newer Optim-insulated leads, out of more than 65,000 leads sold worldwide since 2008.

An externalized conductor occurs when the cable, or conductor, from inside the lead wears through the outer silicone insulation around the lead and appears outside the insulation body. This can be visualized on an X-ray or fluoroscopic image; however, these cables are protected by an additional layer of insulation, and therefore can continue to function properly if an externalization occurs.

For more information: www.sjm.com

Related Content

Boston Scientific, HeartLogic Heart Failure Diagnostic Service, MultiSENSE trial data, AHA Scientific Sessions 2016
News | Heart Failure| November 18, 2016
Boston Scientific recently announced results from the first clinical trial evaluating the performance of the HeartLogic...
atrial fibrillation, warfarin, dementia, Intermountain Medical Center Heart Institute, American Heart Association, AHA Scientific Sessions 2016
News | Atrial Fibrillation| November 15, 2016
Atrial fibrillation patients who use warfarin to lower risk of stroke are at higher risk of developing dementia than...
Medtronic, Claria MRI Quad CRT-D SureScan, FDA approval
Technology | Cardiac Resynchronization Therapy Devices (CRT)| November 15, 2016
Medtronic plc  has received U.S. Food and Drug Administration (FDA) approval for the Claria MRI Quad Cardiac...
catheter ablations, atrial fibrillation, stroke risk, Intermountain Medical Center Heart Institute study, AHA Scientific Sessions, American Heart Association
News | Ablation Systems| November 14, 2016
Atrial fibrillation patients with a prior history of stroke who undergo catheter ablation lower their long-term risk of...
Sponsored Content | Videos | Heart Valve Technology| November 14, 2016
William Abraham, M.D., FACC, discusses advances in heart failure device treatment technologies at the Transcatheter C
genetic testing, sudden cardiac death of teen, Mayo Clinic Proceedings
News | Genetic Testing| November 09, 2016
The recent, sudden death of a 13-year-old boy resulted in more than 20 relatives being incorrectly diagnosed as having...
transcatheter aortic valve replacement, pacemakers, implantation, post-TAVR implantation, worse outcomes, study
News | EP Lab| November 09, 2016
Patients who undergo minimally invasive heart valve replacement, known as transcatheter aortic valve replacement (TAVR...
RFID inventory control in the cath lab, inventory management, cardinal

An example of RFID cabinets in a cath lab. As items are pulled from the cabinet, the inventory control system automatically determines what items were take out and adds them to the patient case. The system can also help locate recalled or expired items, and automatically track on-hand inventory to avoid manual counts.

Feature | Inventory Management| October 28, 2016 | Jean-Claude Saghbini
The healthcare industry’s transition to value-based care leaves no room for waste, and yet we know that inefficiency
Sponsored Content | Videos | Sudden Cardiac Arrest| October 28, 2016
It is critical to educate patients who are at risk of sudden cardiac arrest (SCA), so Rahul Doshi, M.D., director of
Overlay Init