Covidien Multi-function Defibrillation Electrodes Recalled Over Compatibility With Philips FR3 or FRx AED Units

Issue could lead to delay in therapy for users

Covidien announced that it has notified customers of a voluntary Field Safety Alert for certain Medi-Trace Cadence and Kendall Multi-function Defibrillation Electrodes due to a connector compatibility issue with Philips FR3 and FRx defibrillators.

Covidien has received two reports where customers attempted to use a Covidien electrode with a Philips FR3 automated external defibrillator (AED) unit. The mismatch of these devices contributed to a delay in resuscitation and may have contributed to the subsequent death of one patient. 

These electrodes will not connect with Philips FR3 or FRx AED units, and in the case of the use of Covidien defibrillation electrodes with the Philips FR3 AED units, could result in a delay of therapy. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user will not discover the compatibility issue until the AED must be used. This may result in a delay in therapy.  

Covidien alerted customers to the this issue by letter on Sept. 18, 2014, and has revised labeling to clarify that use of these electrodes is incompatible with Philips FR3 and FRx AED units. There are a total of 644,460 electrodes affected by this safety alert. See firm press release for electrodes affected including additional brands.

Philips FR3 and FRx AED units should only be used with the Philips brand electrodes specified in the equipment manual. Customers should review the use of Covidien defibrillation electrodes in their facility to assure that Covidien electrodes are not placed for use with Philips model FR3 or FRx AEDs. 
For further information or to report a problem, please contact Covidien Quality Assurance at 1-800-962-9888, option 8, then extension 2500 between the hours of 8 a.m. and 5 p.m. (eastern); or email [email protected]

Healthcare providers and consumers may report adverse events or quality problems experienced with the use of any Main Street products to the FDA’s MedWatch Adverse Event Reporting program.

For more information: http://www.fda.gov/Safety/MedWatch/SafetyInformation/SafetyAlertsforHumanMedicalProducts/ucm418644.htm

Related Content

AUM Cardiovascular Receives FDA Approval for CADence ECG Device
Technology | ECG| August 08, 2017
AUM Cardiovascular announced it has received clearance from the U.S. Food and Drug Administration (FDA) for CADence, a...
More Than 20 Percent of Low-Risk Patients Receive Annual ECG
News | ECG| July 20, 2017
July 20, 2017 — More than one in five Ontario patients receive an ...
Cardiologs ECG Analysis Platform Receives FDA Clearance
Technology | ECG Monitoring Services| July 13, 2017
Cardiologs Technologies SAS announced that it has received U.S. Food and Drug Administration (FDA) clearance of its...
iRhythm and Stanford Researchers Develop Deep Learning-Based Cardiac Arrhthymia Detection Algorithm
News | Artificial Intelligence| July 11, 2017
iRhythm Technologies Inc. announced a collaboration with the Stanford Machine Learning Group that has resulted in the...
wearable defibrillators prevent sudden cardiac death in pediatric patients
Feature | Sudden Cardiac Arrest| May 11, 2017
May 11, 2017 — A new study shows that the use of a wearable cardioverter defibrillator (WCD) is safe and effective in
Physio-Control Launches HeartSine samaritan PAD 360P Automated External Defibrillator in United States
Technology | Defibrillator Monitors| April 24, 2017
Physio-Control announced April 19 that the company’s HeartSine samaritan PAD 360P (SAM 360P) fully automatic external...
AEDs, automated external defibrillators, requirements, U.S. schools, JACC study
News | Defibrillator Monitors| March 30, 2017
Automated external defibrillators (AEDs) are associated with increased survival of sudden cardiac arrest when installed...
Philips, FDA, class I recall, HeartStart MRx Monitor/Defibrillator
News | Defibrillator Monitors| March 28, 2017
Philips and the U.S. Food and Drug Administration (FDA) announced the company has initiated a recall of its HeartStart...
Zoll, Hospital Wearable Defibrillator, HWD, FDA premarket approval
Technology | Defibrillator Monitors| March 10, 2017
Zoll Medical Corp. announced that the company’s Hospital Wearable Defibrillator (HWD) has been granted premarket...
Sponsored Content | Videos | EP Mapping and Imaging Systems| February 01, 2017
This video, provided by Medtronic, demonstrates the CardioInsight electro-anatomical mapping system.
Overlay Init