Feature | September 13, 2013| Dave Fornell

Patient Killed During Nuclear Imaging Scan

Patient’s death from falling scanner component spurs major recall

The gamma camera portion of SPECT/CT systems (the box-shaped structures mounted to the front of the CT gantry) can weigh well over 1,000 pounds.

Diagnostic imaging is generally considered safe and noninvasive, so it is extremely unusual for a patient to die from injuries received from a scanner. However, this was the case in early June when a patient was killed because a portion of a SPECT/CT scanner fell during the scan at the James J. Peters VA Medical Center in the Bronx, N.Y. 

The weight of nuclear imaging cameras is around 1,300 pounds or more due to it being lead-lined, plus additional lead used in
the construction of the internal collimation system. This significant weight poses design challenges in these systems because of the fatigue on its supports. This is an issue faced by all vendors’ systems. 

In this case, it was a GE Healthcare Infinia Hawkeye 4 system, which prompted the company to initiate a Class I recall of several of its nuclear imaging systems starting in June.  The recall was expanded to most of GE’s SPECT/CT systems in July. 

In its July 3 recall letter, GE said it was informed of the patient’s death on June 5. GE Healthcare responded and worked with the facility and all appropriate government agencies to complete a thorough investigation.

The company was permitted access to the equipment on June 17. GE Healthcare determined the bolts securing the camera to the gantry were loose, thereby stressing the support mechanism and resulting in the incident. The safety concern is related to a potential patient entrapment or crush hazard if the camera falls during a patient exam, the company said.

Soon after the inspection, GE issued its first recall of several of its nuclear imaging systems. However, engineers realized the similarities in the design of support mechanisms across many of GE Healthcare’s SPECT and SPECT/CT imaging systems, so a second recall letter in July expanded the recall to nearly all of GE’s nuclear imaging systems.  The recalls included the Infinia nuclear medicine systems, VG and VG Hawkeye nuclear medicine systems, Helix nuclear medicine systems, Brivo NM615, Discovery NM630, Optima NM/CT640 and the Discovery NM/CT670.

Healthcare facilities were instructed to cease use of the above mentioned nuclear imaging systems until a GE field engineer was able to do a complete inspection of the system and perform any necessary repairs at no cost. If an issue with the support mechanism fasteners was found on a system, the GE field engineer coordinated the replacement of impacted parts in the gantry to ensure the system would operate safely and met specifications.

As of Aug. 1, GE reported it had nearly completed inspections of its SPECT systems installed globally and that systems requiring repair would be returned to service shortly.  

Related Content

ASNC, SNMMI, position statement, myocardial perfusion PET, coronary artery disease
News | PET Imaging| September 01, 2016
September 1, 2016 — The American Society for Nuclear Cardiology (ASNC) and the Society of...
UltraSPECT, Xpress3.Cardiac, nuclear imaging, RWJPE, New Jersey, Robert Wood Johnson Physician Enterprise
News | SPECT Imaging| July 25, 2016
UltraSPECT Inc. announced recently that Robert Wood Johnson Physician Enterprise (RWJPE), a multi-specialty, community-...
ASNC, SNMMI, PET standard, Pet cardiac imaging, PET nuclear imaging, perfusion imaging

An example of a cardiac PET-CT imaging, should a color coded image of the left ventricle to help assess myocardial perfusion to detect ischemia using a Siemens scanner.

News | Nuclear Imaging| July 19, 2016
 
Cell>Point, license agreement, China, United Eastern Pharmaceutical, Oncardia imaging agent
News | Radiopharmaceuticals and Tracers| July 07, 2016
July 7, 2016 — Cell>Point announced in mid-June it has entered into an...
SNMMI, appropriate use criteria, AUC, PAMA, qualified provider-led entity, PLE
News | Clinical Decision Support| June 27, 2016
The Society of Nuclear Medicine and Molecular Imaging (SNMMI) has been named a qualified provider-led entity (PLE)...
Philips, SNMMI 2016, Vereos digital PET/CT, time-of-flight, IntelliSpace 8.0, nuclear imaging
News | Nuclear Imaging| June 14, 2016
Philips announced it would be showcasing a variety of nuclear imaging solutions at the Society of Nuclear Medicine and...
Siemens Healthineers, SNMMI '16, molecular imaging, syngo.via, Biograph Horizon PET/CT, mobile configuration
Technology | Nuclear Imaging| June 09, 2016
At the 2016 Annual Meeting of the Society of Nuclear Medicine and Molecular Imaging (SNMMI), Siemens Healthineers will...
nuclear medicine, global market, radiopharmaceuticals, MEDraysintell, SNMMI 2016
News | Nuclear Imaging| June 08, 2016
MEDraysintell estimates that the global market for nuclear medicine (radiopharmaceuticals) reached $4.3 billion in 2015...
cardiac PET, myocardial perfusion, PET-CT, cardiac perfusion

A PET-CT cardiac perfusion exam from a Siemens Biograph scanner. The black and white areas of the image show the CT imaging of the anatomy. The colored portion shows the PET overlay on the myocardium and is color-coded to show tracer uptake values. This can show areas of the heart muscle where there are perfusion defects cause by infarcts or coronary artery blockages due to a heart attack and help determine the severity of the ischemia.

Feature | PET Imaging| June 03, 2016 | Dave Fornell
Positron emission tomography (PET) is a nuclear imaging technology (also referred to as molecular imaging) that enabl
Medic Vision, SafeCT-29, NEMA XR-29, Smart Dose, computed tomography
Technology | Computed Tomography (CT)| May 31, 2016
Medic Vision Imaging Solutions Ltd. announced the U.S. Food and Drug Administration (FDA) clearance of SafeCT-29 to...
Overlay Init