Feature | September 13, 2013 | Dave Fornell

Patient Killed During Nuclear Imaging Scan

Patient’s death from falling scanner component spurs major recall

The gamma camera portion of SPECT/CT systems (the box-shaped structures mounted to the front of the CT gantry) can weigh well over 1,000 pounds.


Diagnostic imaging is generally considered safe and noninvasive, so it is extremely unusual for a patient to die from injuries received from a scanner. However, this was the case in early June when a patient was killed because a portion of a SPECT/CT scanner fell during the scan at the James J. Peters VA Medical Center in the Bronx, N.Y. 

The weight of nuclear imaging cameras is around 1,300 pounds or more due to it being lead-lined, plus additional lead used in
the construction of the internal collimation system. This significant weight poses design challenges in these systems because of the fatigue on its supports. This is an issue faced by all vendors’ systems. 

In this case, it was a GE Healthcare Infinia Hawkeye 4 system, which prompted the company to initiate a Class I recall of several of its nuclear imaging systems starting in June.  The recall was expanded to most of GE’s SPECT/CT systems in July. 

In its July 3 recall letter, GE said it was informed of the patient’s death on June 5. GE Healthcare responded and worked with the facility and all appropriate government agencies to complete a thorough investigation.

The company was permitted access to the equipment on June 17. GE Healthcare determined the bolts securing the camera to the gantry were loose, thereby stressing the support mechanism and resulting in the incident. The safety concern is related to a potential patient entrapment or crush hazard if the camera falls during a patient exam, the company said.

Soon after the inspection, GE issued its first recall of several of its nuclear imaging systems. However, engineers realized the similarities in the design of support mechanisms across many of GE Healthcare’s SPECT and SPECT/CT imaging systems, so a second recall letter in July expanded the recall to nearly all of GE’s nuclear imaging systems.  The recalls included the Infinia nuclear medicine systems, VG and VG Hawkeye nuclear medicine systems, Helix nuclear medicine systems, Brivo NM615, Discovery NM630, Optima NM/CT640 and the Discovery NM/CT670.

Healthcare facilities were instructed to cease use of the above mentioned nuclear imaging systems until a GE field engineer was able to do a complete inspection of the system and perform any necessary repairs at no cost. If an issue with the support mechanism fasteners was found on a system, the GE field engineer coordinated the replacement of impacted parts in the gantry to ensure the system would operate safely and met specifications.

As of Aug. 1, GE reported it had nearly completed inspections of its SPECT systems installed globally and that systems requiring repair would be returned to service shortly.  


Related Content

News | SPECT-CT

September 12, 2022 — Royal Philips, a global leader in health technology, announced new milestones in the development of ...

Home September 12, 2022
Home
News | SPECT-CT

September 7, 2022 — The American Society of Nuclear Cardiology (ASNC) and three partner societies have come together to ...

Home September 07, 2022
Home
Technology | SPECT-CT

October 18, 2016 — At the 2016 annual meeting of the Radiological Society of North America (RSNA 2016), Siemens ...

Home October 18, 2016
Home
News | SPECT-CT

May 5, 2016 — Doctors and researchers at Barnes-Jewish Hospital and Washington University in St. Louis may be able to ...

Home May 05, 2016
Home
Subscribe Now