News | October 29, 2008

Thoratec Issues Warning For Certain HeartMate II VADs

October 30, 2008 - Thoratec Corp. this week notified healthcare professionals of a worldwide correction of the HeartMate II Left Ventricular Assist System (HM II LVAS), of all serial numbers (having catalog numbers 1355 or 102139), distributed since the beginning of clinical studies in November 2003.

Over time, wear and fatigue of the percutaneous lead connecting the HeartMate II LVAS blood pump with the system controller may result in damage that could interrupt pump function, the company said. This may require reoperation to replace the pump, and potentially result in serious injury or death, Thoratec said. The estimated probability of the need for pump replacement due to percutaneous lead damage is 1.3 percent at 12 months, 6.5 percent at 24 months, and 11.4 percent at 36 months. Healthcare professionals with patients supported by a HeartMate II LVAS should assess the wear and fatigue of the percutaneous lead and provide proper instruction to patients on the management and care of the lead.

For more information: www.fda.gov/medwatch/safety/2008/safety08.htm#HeartMate


Related Content

News | Ventricular Assist Devices (VAD)

October 31, 2023 — Tenaya Therapeutics, Inc., a clinical-stage biotechnology company with a mission to discover, develop ...

Home October 31, 2023
Home
News | Ventricular Assist Devices (VAD)

June 7, 2023 — Magenta Medical Ltd. has announced the initiation of its FDA-approved Early Feasibility Study with the ...

Home June 07, 2023
Home
News | Ventricular Assist Devices (VAD)

March 1, 2023 — Mesoblast Limited, global leader in allogeneic cellular medicines for inflammatory diseases, today ...

Home March 01, 2023
Home
News | Ventricular Assist Devices (VAD)

November 2, 2022 — For decades, left ventricular-assist devices (LVADs) have extended the lives of people whose hearts ...

Home November 02, 2022
Home
News | Ventricular Assist Devices (VAD)

October 18, 2022 — The U.S. Food and Drug Administration (FDA) is providing updates to the FDA website to include ...

Home October 18, 2022
Home
News | Ventricular Assist Devices (VAD)

August 25, 2022 — The U.S. Food and Drug Administration (FDA) issued a statement that Medtronic, Inc. is recalling ...

Home August 25, 2022
Home
News | Ventricular Assist Devices (VAD)

August 5, 2022 — Black people and women with severe heart failure who might be good candidates for surgery to implant a ...

Home August 05, 2022
Home
News | Ventricular Assist Devices (VAD)

June 23, 2022 — Medtronic, Inc. is recalling a single lot of HeartWare HVAD System batteries due to welding defects that ...

Home June 23, 2022
Home
News | Ventricular Assist Devices (VAD)

June 8, 2022 — The U.S. Food and Drug Administration (FDA) has issued a release stating that Medtronic is recalling the ...

Home June 08, 2022
Home
News | Ventricular Assist Devices (VAD)

April 28, 2022 — The U.S. Food and Drug Administration (FDA) is alerting healthcare providers to the possibility that ...

Home April 28, 2022
Home
Subscribe Now