Feature | September 13, 2013| Dave Fornell

Patient Killed During Nuclear Imaging Scan

Patient’s death from falling scanner component spurs major recall

The gamma camera portion of SPECT/CT systems (the box-shaped structures mounted to the front of the CT gantry) can weigh well over 1,000 pounds.

Diagnostic imaging is generally considered safe and noninvasive, so it is extremely unusual for a patient to die from injuries received from a scanner. However, this was the case in early June when a patient was killed because a portion of a SPECT/CT scanner fell during the scan at the James J. Peters VA Medical Center in the Bronx, N.Y. 

The weight of nuclear imaging cameras is around 1,300 pounds or more due to it being lead-lined, plus additional lead used in
the construction of the internal collimation system. This significant weight poses design challenges in these systems because of the fatigue on its supports. This is an issue faced by all vendors’ systems. 

In this case, it was a GE Healthcare Infinia Hawkeye 4 system, which prompted the company to initiate a Class I recall of several of its nuclear imaging systems starting in June.  The recall was expanded to most of GE’s SPECT/CT systems in July. 

In its July 3 recall letter, GE said it was informed of the patient’s death on June 5. GE Healthcare responded and worked with the facility and all appropriate government agencies to complete a thorough investigation.

The company was permitted access to the equipment on June 17. GE Healthcare determined the bolts securing the camera to the gantry were loose, thereby stressing the support mechanism and resulting in the incident. The safety concern is related to a potential patient entrapment or crush hazard if the camera falls during a patient exam, the company said.

Soon after the inspection, GE issued its first recall of several of its nuclear imaging systems. However, engineers realized the similarities in the design of support mechanisms across many of GE Healthcare’s SPECT and SPECT/CT imaging systems, so a second recall letter in July expanded the recall to nearly all of GE’s nuclear imaging systems.  The recalls included the Infinia nuclear medicine systems, VG and VG Hawkeye nuclear medicine systems, Helix nuclear medicine systems, Brivo NM615, Discovery NM630, Optima NM/CT640 and the Discovery NM/CT670.

Healthcare facilities were instructed to cease use of the above mentioned nuclear imaging systems until a GE field engineer was able to do a complete inspection of the system and perform any necessary repairs at no cost. If an issue with the support mechanism fasteners was found on a system, the GE field engineer coordinated the replacement of impacted parts in the gantry to ensure the system would operate safely and met specifications.

As of Aug. 1, GE reported it had nearly completed inspections of its SPECT systems installed globally and that systems requiring repair would be returned to service shortly.  

Related Content

Australian Team Finds New Method for Producing PET Radiotracers in Higher Radiochemical Yields
News | Radiopharmaceuticals and Tracers| April 28, 2017
Researchers at the Australian Nuclear Science and Technology Organisation (ANSTO) have led the development of a new...
Sponsored Content | Videos | Nuclear Imaging| April 28, 2017
David Wolinsky, M.D., director of nuclear cardiology at Cleveland Clinic Florida and past-president of the American S

Physicians will need to use a CMS-certified appropriate use criteria (AUC) clinical decision support software that documents the appropriateness of an imaging order to receive full reimbursement for Medicare patients starting Jan. 1, 2018.

Feature | Cardiac Imaging| April 18, 2017 | Dave Fornell
As part of U.S. healthcare reform efforts, starting Jan.
University of Missouri Research Reactor Files to Start U.S. Production of Medical Isotopes
News | Radiopharmaceuticals and Tracers| April 13, 2017
The University of Missouri Research Reactor (MURR) and its partners Nordion and General Atomics (GA), announced in...
IBA Molecular and Mallinckrodt Nuclear Medicine Merge to Become Curium
News | Radiopharmaceuticals and Tracers| April 11, 2017
April 11, 2017 — IBA Molecular announced that it has merged with previous acquisition Mallinckrodt...
GE Healthcare, HealthTrust, supply agreement, nuclear imaging, radiopharmaceuticals
News | Radiopharmaceuticals and Tracers| April 05, 2017
April 5, 2017 — GE Healthcare has signed an agreement with HealthTrust, a group purchasing organization headquartered
Northwestern Memorial Hospital, MR-PET scanner, first in Illinois, Siemens Biograph mMR
News | PET-MRI| March 31, 2017
March 31, 2017 — Northwestern Memorial Hospital is now home to the Chicago area's first combined magnetic resonance (
PET imaging, atherosclerotic plaque, inflammation, Ga-68-pentixafor, Technishe Universitat Munchen, Germany

Note the high uptake of Ga-68-pentixafor on multi-planar reconstructions in the organs expressing CXCR4 such as the spleen (red arrows) and adrenal glands (yellow arrows), which was nearly completely blocked by the pre-injection of AMD 3100, a potent CXCR4 inhibitor. Strong accumulation of Ga-68-pentixafor was also found in the kidneys (asterisks) reflecting the renal clearance of the tracer. In addition, high, focal activities were detected in the abdominal aorta (red arrowheads) and right carotid artery (orange arrowheads) of atherosclerotic rabbits, whereas no significant signal could be detected in the non-injured left carotid artery (white arrowheads) of atherosclerotic and control rabbits, as well as in the abdominal aorta and right carotid artery of control rabbits. Furthermore, focal activities detected with PET in atherosclerotic plaques of the abdominal aorta and the right carotid artery decreased significantly when the same rabbit was re-imaged after blocking CXCR4 receptors. Image courtesy of Fabien Hyafil, M.D., Ph.D., Department of Nuclear Medicine, Klinikum Rechts der Isar, Technische Universität München, Munich, Germany

News | PET Imaging| March 03, 2017
In the featured article of the March 2017 issue of The Journal of Nuclear Medicine, researchers demonstrate that a new...
Nuclear cardiology, nuclear imaging, radiotracer production, automated radiosynthesis module, myocardial perfusion imaging
Feature | Radiopharmaceuticals and Tracers| March 01, 2017 | Anamika Kumari
Huge portions of the globally produced radiotracers find their origin within geographically centralized, commercial r
GE Healthcare, Rapiscan stress imaging agent, North America, Rapidscan Pharma Solutions Inc.
News | Pharmaceuticals| February 07, 2017
GE Healthcare’s Life Sciences business announced in January that it acquired Rapidscan Pharma Solutions Inc., which has...
Overlay Init