News | Intra-Aortic Balloon Pumps (IABP) | November 05, 2018

Maquet/Datascope Intra-Aortic Balloon Pumps Might Fail on Battery Power

Vendor issues letter to healthcare providers warning of 75 cases of device shut down

he U.S. Food and Drug Administration (FDA) is evaluating recent reports of Getinge's Maquet/Datascope intra-aortic balloon pump (IABP) devices shutting down while running on battery power.

November 5, 2018 — The U.S. Food and Drug Administration (FDA) is evaluating recent reports of Getinge's Maquet/Datascope intra-aortic balloon pump (IABP) devices shutting down while running on battery power. The company issued a letter to providers warning them about this issue Nov. 1, 2018. 

IABPs are used on critically-ill patients in health care facilities, including during transport. The vendor said an interruption in treatment can result in serious patient injury or death.

The Maquet/Datascope IABP is a cardiac assist device placed in the descending aorta, just distal to the left subclavian artery. The device is an electromechanical system used to inflate and deflate intra-aortic balloons, which provides temporary support to the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Getinge manufactures the following Maquet/Datascope IABP devices: Cardiosave (Hybrid and Rescue), CS300 and CS100/CS100i.

Since 2017, the FDA has received more than 75 medical device reports of Maquet/Datascope IABP devices shutting down while running on battery, leading to pump stop and loss of hemodynamic support. Five of these reports described three patient deaths, although the deaths cannot be definitively attributed to the device shutting down. Other reports describe serious patient injury such as a sudden drop in blood pressure or the need for resuscitation. At least six reports indicated that there was no alarm warning before the device shutdown.

The company said the onset of the shutdown as described in the reports ranges from immediately following disconnection of the IABP from AC power (electricity) to within the expected battery run time.

In addition to device shutdown, there are reports describing several battery issues that could lead to patient harm, such as the device not charging the battery, the battery charge indicator not working properly or not showing the correct status of the battery charge, or the battery depleting sooner than expected.

"At this time, the root cause or incidence rate of these IABP devices shutting down while on batteries is not known. Although we are concerned about the device shutdown events associated with Maquet/Datascope IABPs, we recognize that these systems may be the best option for circulatory support for some patients. The FDA wants to ensure you are aware of these device failures that have been observed in patients treated with Maquet/Datascope IABPs and reported to the FDA," the company stated it is letter.

 

Recommendations for Providers

The FDA recommends that healthcare providers should follow each device's operating instructions manual for recommendations on usage, charging, maintenance and storage of the system batteries, since battery run times and discharge cycles vary between IABP models. 

In addition, the FDA recommends that users and servicers of Maquet/Datascope IABP devices should:
   • Ensure the IABP system is plugged into an AC outlet whenever possible during patient use to prevent the battery from depleting.
   • Ensure the IABP system is plugged into an AC outlet when the system is not in use. The batteries should be kept at a full charge even when the IABP is not in use.

When transporting patients within or between facilities, please refer to the system's Operating Instructions Manual for recommendations for portable/battery operation. For example:
   • Prior to portable operation, the battery should be fully charged.
   • Additional charged batteries should be on hand during transport, if applicable for the system. 
   • Ensure the batteries are properly seated in the battery compartment/charger. 
   • Periodically check battery run time and replace batteries as required, as recommended in each system's Operating Instructions Manual. A reduction in run time can occur over a battery's life for reasons such as age, storage temperature and discharge depth. 

Batteries should be replaced:
   • After reaching the maximum number of charge-discharge cycles. 
   • When the battery provides less than the minimum expected run time.
   • When the labeled lifetime of the battery is reached.
   • If the battery is broken, cracked, leaking or damaged. 

The vendor encourages providers to report to the FDA events of IABP devices shutting down while running on batteries, as well as any other battery issues that occur. Voluntary reports can be submitted through MedWatch, the FDA Safety Information and Adverse Event Reporting program. Healthcare personnel employed by facilities subject to FDA's user facility reporting requirements should follow the reporting procedures established by their facilities. Prompt reporting of adverse events can help the FDA identify and better understand the risks associated with medical devices.

When possible, return devices associated with, or suspected to be associated with, any adverse events or device malfunction to the manufacturer for evaluation to help them and the FDA better understand the issue.

The FDA is working with the device manufacturer to better understand these device shutdown events while running on battery for Maquet/Datascope IABPs, and any contributing factors. The FDA will keep the public informed if significant new information becomes available.

Links to Related Maquet IABP FDA Recalls:

Getinge Issues Worldwide Voluntary Correction of Maquet/Getinge Cardiosave Intra-Aortic Balloon Pump

Maquet Datascope Corp./Getinge Group Recalls the Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pumps (IABPs) Due to Possible Malfunction and Failure at High AltitudesGetinge Issues Worldwide Voluntary

Correction of Maquet/Getinge Cardiosave Intra-Aortic Balloon Pump (IABP) For Interruption and/or Inability to Start Therapy at Altitudes above 3,200 Feet/975 Meters (September 20, 2018)

Related Content

Videos | Cath Lab | July 15, 2020
Richard Botto, CVT, RCSA, chief cardiovascular technologist, division of cardiology, cardiac cath lab, offers an over
Videos | Cath Lab | July 13, 2020
The Vieussens’ arterial ring (VAR) is a connection between the conus artery and the left anterior descending (LAD) co
News | Cath Lab | July 06, 2020
July 6, 2020 — Black patients who undergo percutaneous coronary intervention (PCI) are at an increased risk for major
TCT cancels in-person meeting and goes virtual due to COVID-19
News | Cath Lab | May 27, 2020
May 27, 2020 — To ensure the health and safety of all attendees due to the ongoing COVID-19 (SARS-CoV-2) pandemic, th
Patients undergoing cardiac catheterization are traditionally instructed to follow nothing by mouth, or nil per os (NPO), as there are no current standardized fasting protocols, but the CHOWNOW study found patients do not need to fast and will have similar outcomes. #SCAI2020
Feature | Cath Lab | May 18, 2020
May 18, 2020 – Patients undergoing cardiac catheterization are traditionally instructed to follow nothing by mouth, o
Ehtisham Mahmud, M.D., FSCAI, president of the Society for Cardiovascular Angiography and Interventions (SCAI) and chief, Division of Cardiovascular Medicine at UC San Diego Medical Center,
Podcast | Cath Lab | May 13, 2020
This podcast is an interview with Ehtisham Mahmud, M.D., FSCAI, president of the Society for Cardiovascular Angiograp
Nuance Communications Inc. introduced Nuance Cardiovascular CAPD, a new computer-assisted physician documentation (CAPD) solution designed to help cardiologists improve the quality of complex documentation and the accuracy of reimbursement for cardiac catheterization procedures. The Nuance Cardiovascular CAPD solution is available through a partnership with ZHealth for this solution, which is based on patented algorithms built with ZHealth’s interventional documentation and coding expertise.
News | Cath Lab | January 31, 2020
January 29, 2020 – Nuance Communications Inc.