News | June 11, 2009

FDA Calls Class I Recall on Some Medtronic Kappa, Sigma Pacemakers

June 11, 2009 - Medtronic Inc. yesterday said the FDA has classified Medtronic’s previously announced physician advisory about a small subset of Kappa and Sigma series pacemakers, identified at increased susceptibility for separation of bonded interconnect wires from the electronic circuit, as a class I recall.

Notification to FDA, physicians and patients began May 18. As of the date of this news release, more than 95 percent of physicians following affected devices in the U.S. have confirmed receipt of the notification. The FDA’s classification of Medtronic’s Important patient safety Information does not alter the patient management recommendations already provided in the May 2009 letter, and physicians need not take any new action.

Not all Medtronic Kappa and Sigma pacemakers are included in the advisory. Medtronic’s May 2009 communication to physicians involves a small subset of approximately 2 percent of the 1.7 million Kappa and Sigma pacemakers implanted worldwide since 1997. Medtronic has identified an increased susceptibility for separated interconnect wires in a specific subgroup of 10,000 of these devices in the U.S. that remain in patients. Most of the devices in the potentially affected subset have been implanted for five years or longer and may be nearing normal elective replacement time, Medtronic said.

In this limited subset of Kappa and Sigma series pacemakers, the issue can lead to loss of rate response, premature battery depletion, loss of telemetry, or no output, potentially causing patient symptoms such as fainting and lightheadedness, Medtronic said. Patients who experience these symptoms are reminded to seek medical attention immediately.

Patients can determine if their device is included in the identified subset by entering their device serial number at www.KappaSigmaSNList.medtronic.com. Patients whose serial numbers match any of those found on the Web site should contact their physician to determine an appropriate course of action. Patients without Internet access, or who have these models of devices but are unsure if their devices are included in this advisory may also contact Medtronic Patient Services at (800) 551-5544.

After consultation with the company’s Independent Physician Quality Panel, Medtronic’s May 2009 letter recommended that physicians consider device replacement for patients who are both pacemaker dependent and who have been implanted with a device in the affected subsets. Physicians should continue routine follow up in accordance with standard practice for those patients who are not pacemaker dependent.

For more information: www.medtronic.com

Related Content

Ohio State Researchers Prove Human Heart's 'Battery' Has Multiple Backups
News | EP Lab| July 27, 2017
July 27, 2017 — There is good news when it comes to the heart’s sinoatrial node (SAN), the body’s natural...
Medtronic Reactive ATP Therapy Slows Progression of Atrial Fibrillation in Real-World Population
News | Atrial Fibrillation| July 03, 2017
Medtronic recently announced that its Reactive ATP therapy slows the progression of atrial fibrillation (AF) in...
Biosense webster multielectrode balloon RF ablation catheter

The Biosense Webster multi-electrode balloon RF ablation catheter. Each electrode can have varied power settings to avoid damage to underlying tissues like the esophagus. 

Feature | EP Lab| June 22, 2017 | Dave Fornell
Electrophysiology (EP) technology has been advancing rapidly the past few years with new ablation tools to improve...
Pacemakers and Other Cardiac Devices Can Help Solve Forensic Cases
News | Pacemakers| June 20, 2017
Pacemakers and other cardiac devices can help solve forensic cases, according to a study presented at the European...
Sponsored Content | Videos | EP Lab| May 26, 2017
DAIC Editor Dave Fornell takes a tour of some of the most innovative new electrophysiology (EP) technology at the 201
Sponsored Content | Videos | Leads Implantable Devices| May 25, 2017
Bruce Wilkoff, M.D., director of cardiac pacing and tachyarrhythmia devices at Cleveland Clinic, discusses advancemen
Sponsored Content | Videos | Pacemakers| May 23, 2017
Vivek Reddy, M.D., director of cardiac arrhythmia services and professor of medicine, cardiology, Mount Sinai Hospita
micra leadless pacemaker
Feature | Pacemakers| May 16, 2017
May 16, 2017 - The preliminary results for the Medtronic Micra Transcatheter Pacing System (TPS) Post-Approval Regist
Sponsored Content | Videos | Pacemakers| May 16, 2017
This video, provided by Medtronic, demonstrates the implantation of Micra transcatheter pacing system (TPS).
Sponsored Content | Videos | EP Lab| April 10, 2017
A discussion with Heart Rhythm Society (HRS) President Michael Gold, M.D., Ph.D., director of cardiology and associat
Overlay Init