News | Wearable Sensors | January 22, 2018

FDA Issues Safety Communication on Zoll LifeVest 4000 Wearable Cardioverter Defibrillator

Device failure could lead to lack of shock delivery

FDA Issues Safety Communication on Zoll LifeVest 4000 Wearable Cardioverter Defibrillator

January 22, 2018 — The U.S. Food and Drug Administration (FDA) announced last week it is providing information and recommendations regarding the Zoll LifeVest 4000 due to concerns that the device may fail to deliver treatment to the patient if the device is not replaced soon after displaying "Call for service: Device has a problem that may require service. Call Zoll for service, Message Code 102." Failure to contact Zoll and immediately replace the device after Message Code 102 appears on the device screen may result in serious patient harm or death of the patient because the device may fail to deliver therapy appropriately when needed.

Ideally, the LifeVest monitors the patient's heart, delivering a "treatment shock" as needed to restore the patient's heartbeat to a normal rhythm. However, FDA is aware that in certain cases the LifeVest 4000 may not be able to deliver a life-saving treatment shock to a patient due to a fault that prevents the device from charging its high-energy capacitors. During normal operation, this fault prompts the LifeVest to display "Call for service: Device has a problem that may require service. Call Zoll for service, Message Code 102" on the device screen. The "Message Code 102" alert does not explicitly indicate to the patient that the device cannot be used and that the patient should call Zoll immediately.

To date, FDA is aware of one patient death due to the LifeVest's failure to deliver treatment as expected after Message Code 102 was displayed. According to Zoll,  as of Nov. 14, 2017, there were a total of 33,670 devices distributed, with 24,975 devices distributed in the U.S. About 0.1 percent of the total devices distributed have displayed Message Code 102, which should only be remedied by immediate replacement of the device.

The FDA will continue to work with Zoll to monitor LifeVest 4000 devices for any adverse events related to Message Code 102 or a failure to deliver treatment. The agency will also continue to work with Zoll to identify a permanent solution to the Message Code 102 issue, and will keep the public informed as new information becomes available.

The Zoll LifeVest 4000 is a wearable defibrillator used to treat life-threatening arrhythmias in adults and children who are at risk for sudden cardiac arrest and are not candidates for, or refuse, an implantable defibrillator. The LifeVest continuously monitors the patient's heart and, if a fast, life-threatening heart rhythm is detected, the device delivers a treatment (i.e. a shock) to restore the patient to a normal heart rhythm.

The FDA’s recommended actions for healthcare providers include:

  • Train patients prescribed use of the Zoll LifeVest 4000 on how to identify the "Call for service, Message Code 102" alert. Message Code 102 may first display as a gong alert, and then subsequently at every power-up until the device is returned to Zoll for servicing. The alert will read "Call for service: Device has a problem that may require service. Call Zoll for service, Message Code 102" and will allow the patient to tap "OK" to return to normal operation;
  • Inform patients to respond to the "Call for service, Message Code 102" alert on the LifeVest 4000 screen by contacting Zoll immediately. Message Code 102 may mean that the LifeVest 4000 has a severe service problem and cannot be used because the device may not appropriately deliver life-saving treatment if needed. Zoll will replace the patient's device within 24 hours if the device displays Message Code 102. Zoll's Customer Service can be reached 24 hours a day, and 7 days a week at 1-800-543-3267; and
  • Reinforce initial LifeVest 4000 training, explaining that the device can display messages on the screen and play audio alerts that require the patient to take action. If Message Code 102 does not appear on the device screen, the patient should continue to use the LifeVest 4000 as prescribed, and read the display for gong alerts and follow the instructions on the screen (refer to Section 5 in the Patient Manual).

Recommendations for patients and caregivers include:

  • Contact Zoll immediately if "Call for service: Device has a problem that may require service. Call Zoll for service, Message Code 102" appears on your LifeVest 4000 screen. Message Code 102 may first display as a gong alert, and then subsequently at every power-up until the device is returned to Zoll for servicing. The Message Code 102 alert will also allow you to tap "OK" to return to normal operation;
    • However, the "Call for service, Message Code 102" alert may mean that the LifeVest 4000 has a severe service problem and cannot be used because the device may not appropriately deliver life-saving treatment if needed.
    • Zoll will replace your device within 24 hours if the device displays Message Code 102. Zoll's Customer Service can be reached 24 hours a day, and 7 days a week at: 1-800-543-3267.
  • Continue to use your LifeVest 4000 as prescribed by your physician and as instructed during your initial training if the "Call for service, Message Code 102" alert is NOT displayed on your device screen. You should continue to read the display for gong alerts and follow the instructions on the screen (refer to Section 5 in the Patient Manual).

Healthcare professionals and patients are encouraged to report adverse events or side effects related to the use of these products to the FDA's MedWatch Safety Information and Adverse Event Reporting Program:

  • Complete and submit the report Online: www.fda.gov/MedWatch/report
  • Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178

For more information: www.fda.gov

 

Related Content

Rice University Students Improve AED Operation With Needle Pad

The automated external defibrillator add-on developed by Rice University students is a pad of needles that break through the skin's natural impedance barrier to deliver an electric shock to the heart of a patient in cardiac arrest. (Credit: Jeff Fitlow/Rice University)

News | Defibrillator Monitors | April 16, 2018
The skin presents a formidable barrier to life-saving defibrillators, but a team of students from Rice University...
Zoll Canada Equipping Province of Québec Paramedic Services with X Series Monitor/Defibrillators
News | Defibrillator Monitors | October 16, 2017
October 16, 2017 — Zoll Canada, a subsidiary of Zoll Medical Corp., announced it has won the tender to equip all ambu
Physio-Control Launches HeartSine samaritan PAD 360P Automated External Defibrillator in United States
Technology | Defibrillator Monitors | April 24, 2017
Physio-Control announced April 19 that the company’s HeartSine samaritan PAD 360P (SAM 360P) fully automatic external...
AEDs, automated external defibrillators, requirements, U.S. schools, JACC study
News | Defibrillator Monitors | March 30, 2017
Automated external defibrillators (AEDs) are associated with increased survival of sudden cardiac arrest when installed...
Philips, FDA, class I recall, HeartStart MRx Monitor/Defibrillator
News | Defibrillator Monitors | March 28, 2017
Philips and the U.S. Food and Drug Administration (FDA) announced the company has initiated a recall of its HeartStart...
Zoll, Hospital Wearable Defibrillator, HWD, FDA premarket approval
Technology | Defibrillator Monitors | March 10, 2017
Zoll Medical Corp. announced that the company’s Hospital Wearable Defibrillator (HWD) has been granted premarket...
Zoll LifeVest wearable defibrillator, WEARIT-II Registry results, CardioStim EuroPace 2016
News | Defibrillator Monitors | June 21, 2016
Zoll Medical Corp. announced that patients experience a high one-year survival rate following use of the LifeVest...
defibrillator-monitors, defibrillator monitor

Physio-Control's LifePAK15

Feature | Defibrillator Monitors | February 03, 2014 | Dave Fornell
For U.S. hospitals and emergency medical services (EMS) looking to upgrade their aging defibrillator-monitors, new...
Overlay Init