News | April 14, 2009

Arrow Voluntarily Recalls Intra Aortic Balloon Pump Catheters

April 14, 2009 - The voluntary medical device recall initiated on Feb. 2, 2009 by Arrow International’s Cardiac Care Division involves volume connectors for its 30cc, 40cc and 50cc Intra-aortic balloon pump (IABP) catheters as a class 1 recall.

The FDA defines class I recalls as, “a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.”

The recall is being conducted because a fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 cc or 5 cc, rather than the appropriate 30, 40, or 50 cc volume. In patients exhibiting moderate to severe myocardial ischemia or low perfusion states, prolonged exposure to the low default volume may fail to decrease ischemia and increase perfusion, leading to organ injury or infarct and may result in patient death. Prolonged exposure to a device that is inflated to less than 2/3 full could also result in thrombus formation on the IABP and possible subsequent systemic or cerebral thromboembolism.

Instructions contained in the operator’s manual fully describe the necessary warnings and precautions that will identify the issue. The recall notice restates these instructions and references the sections in the owner’s manual. If these instructions are followed, the issue for which the product was recalled can be discovered immediately, and the product can be removed and replaced.

Arrow International notified both domestic and foreign hospitals and distributors via an urgent medical device recall letter dated Feb. 2, 2009 that the company had become aware that the blue connector for the 40cc IABP was not properly recognized by the Arrow IABP system. This recall involved the retrieval of unused product, issuance of mitigation instructions for patients and/or facilities in critical need, and the replacement of pump tubing assemblies.

The products affected are: 30/40/50 CC IAB Catheters, with product codes: IAB-04830-U, IAB-04840-U, IAB-05830-LWS, IAB05830-U, IAB-05840-U, IAB-05840-LWS, IAB06830-U, IAB06840-U, IAB-S730C, IAB-S840C, IAB-R950-U, IAK-02692, IAK-02693, and IAK-02691.

A list of products and lot numbers affected by this recall as well as the original recall notice can be found at Arrow’s website.

For more information: http://www.arrowintl.com/iab_recall.asp

Related Content

News | Drug-Eluting Balloons| August 15, 2017
Surmodics Inc. announced receipt of an investigational device exemption (IDE) from the U.S. Food and Drug...
Spectranetics Announces FDA Approval of Stellarex Drug-Coated Balloon
Technology | Drug-Eluting Balloons| July 26, 2017
The Spectranetics Corp. announced receipt of U.S. Food and Drug Administration (FDA) pre-market approval (PMA) of the...
IABP recall for Maquet, Datascope intra-aortic balloon pump CS100 and CS300

The Datascope CS300 IABP console.

News | Hemodynamic Support Devices| June 20, 2017
June 20, 2017 — The U.S.
News | June 12, 2017
June 12, 2017 – Med Alliance announced completion on schedule of patient enrollment in the first-in-man (FIM) study o
Abbott recalls its NC Balloon catheters
News | Balloon Catheter| May 22, 2017
May 22, 2017 — Abbott Vascular has initiated a voluntary recall of specific lots of three catheters due to 19 reports
Sponsored Content | Videos | EP Lab| May 17, 2017
This video, provided by Spectranetics, demonstrates how to deploy the Bridge Occlusion Balloon used to seal accidenta
Contego Medical Receives CE Mark for Vanguard IEP Peripheral Balloon Angioplasty System
News | Peripheral Artery Disease (PAD)| May 17, 2017
Contego Medical LLC announced recently that it has received CE Marking of its Vanguard IEP Peripheral Balloon...
Technology | Intra-Aortic Balloon Pumps (IABP)| May 02, 2017
May 2, 2017 — The U.S. Food and Drug Administration (FDA) granted 510(k) clearance for the Teleflex Inc.
Twelve-Month Data Positive for Ranger Paclitaxel-Coated Balloon Catheter
News | Drug-Eluting Balloons| April 26, 2017
Boston Scientific announced results from the RANGER SFA trial for the Ranger Paclitaxel-Coated PTA Balloon Catheter at...
Medtronic, IN.PACT Admiral DCB, drug-coated balloon, below-the-knee, European BTK IN.PACT clinical study, peripheral artery disease, PAD, launch
News | Peripheral Artery Disease (PAD)| March 09, 2017
Medtronic plc announced the launch of the IN.PACT BTK study to evaluate the effectiveness of a drug-coated balloon (DCB...