News | August 03, 2012

St. Jude Medical Announces Initial Findings From Riata Lead Evaluation Study

Study results indicate that externalized conductors occur less frequently in thinner leads

August 3, 2012 — St. Jude Medical recently announced initial findings from the Riata lead evaluation study. The study’s phase I results found that externalized conductors occurred in 9.3 percent of the smaller-diameter Riata ST 7 French leads in the study, and in 24 percent of the larger-diameter Riata 8 French leads.

The Riata lead evaluation study is an international, multicenter, prospective study of Riata and Riata ST silicone defibrillation leads. It enrolled 724 patients at 20 sites in the United States and Canada. An additional 51 patients were enrolled at three sites in Japan. Those results are awaiting final adjudication and will be reported at a later date.

The study will continue to evaluate the performance of Riata leads—both with and without externalized conductors—over the next two years to determine how they function over time, which may help further inform patient management considerations. In other clinical studies, the majority of leads with an externalized conductor continued to function normally.

“The rates of externalized conductors in this study are consistent with other published studies involving fluoroscopic screening of patients with Riata leads. This study also reinforces that externalized conductors are more prevalent in larger-diameter Riata leads,” said Mark Carlson, chief medical officer and senior vice president of clinical affairs for the St. Jude Medical Cardiac Rhythm Management Division. “We are now focused on collecting longer-term data in the second phase of this study. Our goal is to provide physicians timely and relevant clinical data to best support their patient management decisions.”

An experienced physician panel adjudicated fluoroscopic images using objective criteria for the presence or absence of externalized conductors in enrolled patients’ leads. The St. Jude Medical independent Leads Medical Advisory Board has reviewed the results and recommends no changes to existing patient management recommendations.

Based on an analysis of products returned to St. Jude Medical, externalized conductors in Riata leads are most frequently associated with inside-out abrasion, which occurs when the cables housed within the outer insulation of the lead work their way outside of the lead body as a result of movement against the insulation. External, or outside-in, abrasion is a known cause of failure across all cardiac leads in the industry, and is different from the inside-out abrasion seen with externalized conductors in some Riata leads.

Externalized conductors can be visualized through X-ray or fluoroscopy; however, it is important to note that externalized conductors may be visualized without any associated clinical or device-related observations. In the majority of reported cases, a lead with an externalized conductor continued to function normally.

For more information: www.riatacommunication.com

Related Content

An implanted ICD showing its three venous leads. These multiple CRT leads can cause issues when they need to be replaced and are abandoned with new leads put over them in the SVC, which may require lead extraction.

An implanted ICD showing its three leads in the venous system. 

Feature | EP Lab | January 13, 2019 | Dave Fornell, Editor
To extract or abandon broken or infected implantable, venous...
Videos | EP Lab | January 03, 2019
This is a quick video tour of one of the dedicated...
HHS and NASA Team Up to Explore Health on Earth and in Outer Space

Image courtesy of NASA

News | EP Lab | December 14, 2018
The Department of Health and Human Services (HHS) recently entered into an interagency agreement with the National...
Technology Improvements Driving Electrophysiology Market Growth

Navik 3D Cardiac Mapping System image courtesy of APN Health.

News | EP Lab | December 10, 2018
December 10, 2018 — Electrophysiology, whil
Vagal Maneuvers With Supraventricular Tachycardia

Image courtesy of ACLS Certification Institute

Feature | EP Lab | December 05, 2018
Supraventricular tachycardia (SVT) is a common heart abnormality that presents as a fast heart rate. SVT is a generic...
Societies Publish New Guidance for Treatment of Slow, Irregular Heartbeats
News | EP Lab | November 09, 2018
The American College of Cardiology (ACC), the American Heart Association (AHA) and the Heart Rhythm Society (HRS)...
A view of the EPD Solutions catheter ablation system image guidance. It displays the catheter within a pre-acquired 3D segmented CT image. The D700 system provides a real-time lesion assessment tool that predicts transmurality and permanency of lesions, pre- and immediately post-ablation.

A view of the EPD Solutions catheter ablation system image guidance. It displays the catheter within a pre-acquired 3D segmented CT image. The D700 system provides a real-time lesion assessment tool that predicts transmurality and permanency of lesions, pre- and immediately post-ablation.

News | EP Lab | June 05, 2018
June 5, 2018 — Philips Healthcare has signed an agreement to acquire EPD Solutions, a provider of image-guidance in c
A recent study shows the Baylis NRG radiofrequency (RF) Transseptal puncture catheter has a lower incidence of embolism in EP cases.
News | EP Lab | May 21, 2018
May 21, 2018 — A recent study published in Heart and Vessels has found that the use of the Baylis Medical NR
Myocarditis is an Under-recognized Etiology of Symptomatic Premature Ventricular Arrhythmia (PVCs). #HRS #HRS2018
News | EP Lab | May 18, 2018
May 18, 2018 — A significant number of patients with symptomatic premature ventricular contractions (PVCs) have under
Overlay Init