News | June 25, 2012

St. Jude Medical Says Single Durata Lead Failure Caused by External Abrasion

Analysis of lead from Florida hospital shows no inside-out abrasion in Durata product

June 25, 2012 — St. Jude Medical Inc. today commented on the Durata lead failure reported in the MAUDE database report submitted to the U.S. Food and Drug Administration (FDA) May 2.

Through the company's investigation of the MAUDE database report and information provided to the company by the FDA, including model number, implant and event dates, St. Jude Medical identified a single Durata lead that matches the available information. The FDA has confirmed that the serial number of this lead matches the serial number from the MAUDE report.

"Our analysis of the lead indicates it was damaged due to external abrasion, which is not the same as the inside-out abrasion that has been seen in our previous generation Riata leads," the company said. "Simultaneous to this announcement, the company has submitted a medical device report (MDR) to the FDA with our additional findings."

The identified lead was explanted at a Florida hospital and was provided to St. Jude with static and dynamic fluoroscopy and photographs of the extracted lead. The hospital also allowed the company to inspect the lead at the hospital. 

"Through our investigation, we have identified that the patient had an additional defibrillation lead that had been capped (and was no longer being used). In the fluoroscopic images, the two leads cross in the region mentioned in the filed MAUDE report," the company said. "Based upon physical examination, our analysis indicates the damage to the Durata lead is consistent with external abrasion from contact with a calcified, or hardened, heart valve or possibly from lead-to-lead contact. External abrasion is a known cause of failure across all cardiac leads in the industry, which is different from the inside-out abrasion seen with externalized conductors observed in some Riata leads."

"We recognize the importance of providing physicians with up-to-date and accurate information in a timely and responsible manner so that they can make informed patient care decisions. This case highlights the importance of appropriate analysis in determining the root cause of lead damage. It is our desire to work closely with physicians to understand the circumstances surrounding a device malfunction. We encourage physicians to report any case of lead failure to the manufacturer for further inspection and analysis to best ensure we are able to validate and communicate information in the interest of patient safety," St. Jude Medical said in its statement.

For more information: www.sjm.com

Related Content

Lexington Begins HeartSentry Clinical Trial
News | Clinical Study | February 20, 2018
February 20, 2018 – Lexington Biosciences, Inc., a development-stage medical device company, announced the commenceme
Endologix Completes Patient Enrollment in the ELEVATE IDE Clinical Study
News | Clinical Study | February 06, 2018
February 6, 2018 – Endologix, a developer and marketer of treatments for aortic disorders, announced the completion o
12-Month Results from Veryan Medical's MIMICS-2 IDE Study Presented at LINC
News | Clinical Study | February 01, 2018
February 1, 2018 – Thomas Zeller (Bad Krozingen, Germany) presented the 12-month results from Veryan Medical’s MIMICS
LimFlow Completes U.S. Feasibility Study Enrollment, Receives FDA Device Status
News | Clinical Study | February 01, 2018
February 1, 2018 –  LimFlow SA, developer of minimally-inv
ESC 2017 late breaking trial hot line study presentations.
News | Clinical Study | September 12, 2017
September 12, 2017 – The European Society of Cardiology (ESC) Congress 2017 includes several Hot Line Late-breaking C
U.K., NHS studies, weekend effect, hospital admission, atrial fibrillation, heart failure
News | Clinical Study | June 28, 2016
New research shows patients admitted to National Health Service (NHS) hospitals in the United Kingdom for atrial...
stroke risk
News | Clinical Study | August 28, 2015
Most people assume strokes only happen to octogenarians, but recent evidence suggests that survivors of childhood can
Overlay Init