Feature | May 04, 2009| Bjørn Egil Løndalen, M.D., Unilabs AS

Lowering Radiation Dose in Selected Patients Undergoing Thoracic CT

A comparison of the original and Clarity-enhanced CT image from Philips Brilliance 64, showing a 3 mm contrast-enhanced thorax slice at 133 mAs, 120 KVP.

A comparison of a noncontrast, 3 mm slice CT image from a Philips Brilliance 64.


In an effort to lower the radiation doses patients receive from CT scans, Unilabs Röntgen Majorstuen in Oslo, Norway, installed dose-lowering enhancement software on a trial basis between May and November 2008. In selected cases of thoracic CT, the hospital was able to reduce up to 70 percent of the dose as compared to full-dose examinations.

The facility used the Clarity CT Solution software made by Saphenia. The system was connected to the facility’s Philips Brilliance 64 CT and Siemens PACS. Data acquired from the scanner were reconstructed on the scanner before post-processing on the Clarity server. From there DICOM images were transferred to the PACS and Philips workstation. In addition, unprocessed raw data was transferred to the workstation for comparison and secondary reconstruction purposes.

Unilabs chose only to implement the system for low-dose noncontrast thoracic CT examinations during the trial period. Indications for referral included dyspnea, chest pain, possible pneumonia, suspected interstitial lung disease, and skeletal abnormalities or possible finding on a chest X-ray.

There were 35 patients examined under these criteria. Most underwent noncontrast, low-dose CT from the jugulum to the diaphragm at 120 kV, 80 mA. In the last 16 patients the dose was further reduced by lowering mA by an average of 20 percent, some as low as 30 mA. In addition the radiologists included a control group of 10 patients who, for various reasons, had to undergo a full-dose CT of the chest. The images were reconstructed in 3/2 before transfer to the Clarity server.

The series of images were examined by one of the three staff radiologists and reported as any other CT scan. Image quality and clinical value were also evaluated.

The mean radiation dose in the 35 low-dose patients was 2.41 mSv. The mean dose in the 10 full-dose patients was 5.29 mSv, while the national average dose in Norway is 10 mSv (1). Use of the software contributed to a 53.3 percent and 74.1 percent reduction respectively. If only the last 16 patients were considered, the average dose was 1.55 mSv. This translates into an 84 percent reduction from average dose for Norway and a 70 percent reduction from Unilabs’ average full-dose scans. Image quality and diagnostic value was considered good or very good by all reporting radiologists.


(1) Norwegian Radiation Protection Authority http://www.nrpa.no/archive/Internett/Publikasjoner/Veiledere/Veileder5b.pdf, table 6.2


For more information: www.claritysolutions.org, www.unilabs.no

Related Content

Videos | Computed Tomography (CT) | July 17, 2018
An interview with Patrick Serruys, M.D., Ph.D., Imperial College London, principal investigator of the SYNTAX III Tri
Zebra Medical Vision Announces FDA 510(k) Clearance of Coronary Calcium Algorithm
Technology | Computed Tomography (CT) | July 12, 2018
Zebra Medical Vision has received 510(k) clearance from the U.S. Food and Drug Administration (FDA) for its Coronary...
IAC Awards First CT Accreditation for a Mobile Stroke Unit
News | Computed Tomography (CT) | July 09, 2018
The Intersocietal Accreditation Commission (IAC) recently recognized the University of Tennessee Health Science Center...
FFR-CT may soon replace invasive angiography for coronary lesion assessment. #EuroPCR 

Anexample of a clinical case of CT-FFR, which can derive the FFR numbers for their entire coronary tree using a noninvasive CT scan.

Feature | Computed Tomography (CT) | May 29, 2018
May 29, 2018 — Results from the innovative SYNTAX III Revolution Trial [1] underline the effectiveness of evolving no
James Min Named Editor-in-Chief of Journal of Cardiovascular Computed Tomography
News | Computed Tomography (CT) | May 01, 2018
James K. Min, M.D., FSCCT, has been selected for a five-year term as the new editor-in-chief of the Journal of...
FDA Clears Siemens' Somatom go.All, go.Top CT Scanners
Technology | Computed Tomography (CT) | April 18, 2018
April 18, 2018 — The U.S.
Canon Medical Systems' Aquilion Precision CT Receives FDA Clearance
Technology | Computed Tomography (CT) | April 13, 2018
Canon Medical Systems USA Inc. this week received U.S. Food and Drug Administration (FDA) clearance for the Aquilion...
FDA Clears Siemens Somatom Edge Plus CT System
Technology | Computed Tomography (CT) | April 04, 2018
April 4, 2018 — The U.S.
Canon Medical Systems Highlights Aquilion One/Genesis CT Stroke Suite at SIR 2018
News | Computed Tomography (CT) | March 29, 2018
March 29, 2018 — The Aquilion One/Genesis Edition...
CT Shows Enlarged Aortas in Former Pro Football Players

3-D rendering from a cardiac CT dataset demonstrating mild dilation of the ascending aorta. Image courtesy of Christopher Maroules, M.D.

News | Computed Tomography (CT) | January 12, 2018
Former National Football League (NFL) players are more likely to have enlarged aortas, a condition that may put them at...
Overlay Init