Feature | May 04, 2009| Bjørn Egil Løndalen, M.D., Unilabs AS

Lowering Radiation Dose in Selected Patients Undergoing Thoracic CT

A comparison of the original and Clarity-enhanced CT image from Philips Brilliance 64, showing a 3 mm contrast-enhanced thorax slice at 133 mAs, 120 KVP.

A comparison of a noncontrast, 3 mm slice CT image from a Philips Brilliance 64.


In an effort to lower the radiation doses patients receive from CT scans, Unilabs Röntgen Majorstuen in Oslo, Norway, installed dose-lowering enhancement software on a trial basis between May and November 2008. In selected cases of thoracic CT, the hospital was able to reduce up to 70 percent of the dose as compared to full-dose examinations.

The facility used the Clarity CT Solution software made by Saphenia. The system was connected to the facility’s Philips Brilliance 64 CT and Siemens PACS. Data acquired from the scanner were reconstructed on the scanner before post-processing on the Clarity server. From there DICOM images were transferred to the PACS and Philips workstation. In addition, unprocessed raw data was transferred to the workstation for comparison and secondary reconstruction purposes.

Unilabs chose only to implement the system for low-dose noncontrast thoracic CT examinations during the trial period. Indications for referral included dyspnea, chest pain, possible pneumonia, suspected interstitial lung disease, and skeletal abnormalities or possible finding on a chest X-ray.

There were 35 patients examined under these criteria. Most underwent noncontrast, low-dose CT from the jugulum to the diaphragm at 120 kV, 80 mA. In the last 16 patients the dose was further reduced by lowering mA by an average of 20 percent, some as low as 30 mA. In addition the radiologists included a control group of 10 patients who, for various reasons, had to undergo a full-dose CT of the chest. The images were reconstructed in 3/2 before transfer to the Clarity server.

The series of images were examined by one of the three staff radiologists and reported as any other CT scan. Image quality and clinical value were also evaluated.

The mean radiation dose in the 35 low-dose patients was 2.41 mSv. The mean dose in the 10 full-dose patients was 5.29 mSv, while the national average dose in Norway is 10 mSv (1). Use of the software contributed to a 53.3 percent and 74.1 percent reduction respectively. If only the last 16 patients were considered, the average dose was 1.55 mSv. This translates into an 84 percent reduction from average dose for Norway and a 70 percent reduction from Unilabs’ average full-dose scans. Image quality and diagnostic value was considered good or very good by all reporting radiologists.


(1) Norwegian Radiation Protection Authority http://www.nrpa.no/archive/Internett/Publikasjoner/Veiledere/Veileder5b.pdf, table 6.2


For more information: www.claritysolutions.org, www.unilabs.no

Related Content

CT Shows Enlarged Aortas in Former Pro Football Players

3-D rendering from a cardiac CT dataset demonstrating mild dilation of the ascending aorta. Image courtesy of Christopher Maroules, M.D.

News | Computed Tomography (CT)| January 12, 2018
Former National Football League (NFL) players are more likely to have enlarged aortas, a condition that may put them at...
Siemens Healthineers Strengthens CT Portfolio With Four New Systems at RSNA 2017

The Somatom Force with the new FAST 3D Camera

Technology | Computed Tomography (CT)| December 14, 2017
December 14, 2017 — Siemens Healthineers introduced four new...
Toshiba Medical Introduces Aquilion Prime SP CT System at RSNA 2017
Technology | Computed Tomography (CT)| December 13, 2017
December 13, 2017 — Toshiba Medical, a Canon Group company, introduced its new Aquilion Prime SP...
Philips Debuts IQon Elite Spectral CT Scanner at RSNA 2017
Technology | Computed Tomography (CT)| December 11, 2017
At the 2017 Radiological Society of North America (RSNA) Annual Meeting, Philips unveiled the IQon Elite Spectral CT,...
Toshiba Highlights Ultra-High Resolution CT at RSNA 2017
News | Computed Tomography (CT)| December 06, 2017
Toshiba Medical, a Canon Group company, showcased the Aquilion Precision, what it calls the world’s first ultra-high...
Samsung Unveils Mobile CT OmniTom at RSNA 2017
Technology | Computed Tomography (CT)| November 26, 2017
Samsung Electronics debuted its OmniTom mobile 16-slice computed tomography (CT) scanner at the Radiological Society of...
Siemens Healthineers Announces First U.S. Installs of Somatom go.Up CT System
News | Computed Tomography (CT)| November 15, 2017
November 15, 2017 — Center for Diagnostic Imaging (CDI), one of the nation’s largest providers of diagnostic imaging
Medis Releases QAngio CT v3.1
Technology | Computed Tomography (CT)| October 09, 2017
Medis has released a new version of its QAngio CT (computed tomography), which can now be launched from the Medis Suite...
Hitachi Supria True64 CT Receives FDA Clearance
Technology | Computed Tomography (CT)| September 15, 2017
Hitachi Healthcare Americas Inc. announced it has attained U.S. Food and Drug Administration (FDA) 510(k) clearance to...
Orange County, Calif. Hospital Adopts Siemens Somatom Force CT for Cardiac Imaging
News | Computed Tomography (CT)| September 12, 2017
Hoag Memorial Hospital Presbyterian recently became the first hospital in Orange County, Calif., to install the Siemens...
Overlay Init