Feature | May 04, 2009| Bjørn Egil Løndalen, M.D., Unilabs AS

Lowering Radiation Dose in Selected Patients Undergoing Thoracic CT

A comparison of the original and Clarity-enhanced CT image from Philips Brilliance 64, showing a 3 mm contrast-enhanced thorax slice at 133 mAs, 120 KVP.

A comparison of a noncontrast, 3 mm slice CT image from a Philips Brilliance 64.


In an effort to lower the radiation doses patients receive from CT scans, Unilabs Röntgen Majorstuen in Oslo, Norway, installed dose-lowering enhancement software on a trial basis between May and November 2008. In selected cases of thoracic CT, the hospital was able to reduce up to 70 percent of the dose as compared to full-dose examinations.

The facility used the Clarity CT Solution software made by Saphenia. The system was connected to the facility’s Philips Brilliance 64 CT and Siemens PACS. Data acquired from the scanner were reconstructed on the scanner before post-processing on the Clarity server. From there DICOM images were transferred to the PACS and Philips workstation. In addition, unprocessed raw data was transferred to the workstation for comparison and secondary reconstruction purposes.

Unilabs chose only to implement the system for low-dose noncontrast thoracic CT examinations during the trial period. Indications for referral included dyspnea, chest pain, possible pneumonia, suspected interstitial lung disease, and skeletal abnormalities or possible finding on a chest X-ray.

There were 35 patients examined under these criteria. Most underwent noncontrast, low-dose CT from the jugulum to the diaphragm at 120 kV, 80 mA. In the last 16 patients the dose was further reduced by lowering mA by an average of 20 percent, some as low as 30 mA. In addition the radiologists included a control group of 10 patients who, for various reasons, had to undergo a full-dose CT of the chest. The images were reconstructed in 3/2 before transfer to the Clarity server.

The series of images were examined by one of the three staff radiologists and reported as any other CT scan. Image quality and clinical value were also evaluated.

The mean radiation dose in the 35 low-dose patients was 2.41 mSv. The mean dose in the 10 full-dose patients was 5.29 mSv, while the national average dose in Norway is 10 mSv (1). Use of the software contributed to a 53.3 percent and 74.1 percent reduction respectively. If only the last 16 patients were considered, the average dose was 1.55 mSv. This translates into an 84 percent reduction from average dose for Norway and a 70 percent reduction from Unilabs’ average full-dose scans. Image quality and diagnostic value was considered good or very good by all reporting radiologists.


(1) Norwegian Radiation Protection Authority http://www.nrpa.no/archive/Internett/Publikasjoner/Veiledere/Veileder5b.pdf, table 6.2


For more information: www.claritysolutions.org, www.unilabs.no

Related Content

St. Cloud Hospital Installs Toshiba Aquilion One Genesis CT for Emergency Department
News | Computed Tomography (CT)| August 23, 2017
Central Minnesota residents now have access to advanced computed tomography (CT) technology that is safe and fast at St...
Houston Methodist Hospital Enters Multi-Year Technology and Research Agreement With Siemens Healthineers
News | Cardiac Imaging| August 17, 2017
Houston Methodist Hospital and Siemens Healthineers have entered into a multi-year agreement to bring cutting-edge...
CDN to Integrate Advanced Cardiac Imaging Tools From DiA Imaging Analysis
Technology | August 10, 2017
August 10, 2017 — CDN recently announced a new partnership agreement with DiA Imaging Analysis Ltd., makers of next-g
Four Blue Cross Blue Shield Companies Issue Positive Medical Policies on HeartFlow FFRct Analysis
News | FFR Catheters| August 09, 2017
HeartFlow Inc. announced that four Blue Cross Blue Shield companies have each issued a positive medical policy for the...
3D printing of the heart and coronary artery tree from a patient's CT scan.
Sponsored Content | Webinar | 3-D Printing| August 09, 2017
Learn how 3-D printing empowers medical device manufacturer Medtronic to bring products to market faster, develop bet
Virtual Phantoms Inc. Releases VirtualDose-IR
Technology | Radiation Dose Management| August 07, 2017
Virtual Phantoms Inc. announced the release of VirtualDoseIR, a tool for assessing organ dose from interventional...
Moffitt Cancer Center Enhances Patient Care with Toshiba Medical's Infinix-i 4-D
News | Interventional Radiology| August 03, 2017
Cancer patients at Moffitt Cancer Center in Tampa, Fla., now have access to advanced diagnostic imaging for fast and...
New York Hospital Finds Significant Cost Savings With Toshiba's Aquilion One CT
News | Computed Tomography (CT)| July 25, 2017
July 25, 2017 — In five years, Kaleida Health’s Stroke Care Center (SCC) at the Gates Vascular Institute in Buffalo,
Samsung Introduces FDA-Cleared BodyTom Elite CT Scanner
Technology | Computed Tomography (CT)| July 24, 2017
Samsung announced U.S. Food and Drug Administration (FDA) approval of the BodyTom Elite, an upgraded version of its...
Sponsored Content | Videos | CT Angiography (CTA)| July 21, 2017
DAIC and ITN Editor Dave Fornell discusses some of the most innovative new computed tomography (CT) technology and tr
Overlay Init