News | February 24, 2010

320-Slice CT Reduces Radiation Exposure by 90 Percent

February 24, 2010 - Researchers found that when using standard 64-detector row helical scanning as a benchmark, the effective radiation dose was reduced by 91 percent from 35.4 millisieverts (mSv) to 4.4 mSv using optimized 320-detector row volume scanning.

Researchers from Columbia University and the National Heart, Lung and Blood Institute conducted the study, published in the March issue of Radiology, in which they performed a coronary CT angiography exam, and determined that an imaging exam of the heart using the latest generation of CT technology exposes patients to as much as 91 percent less radiation than standard helical CT scanning.

Many coronary CT angiography exams are conducted on 64-detector row CT scanners, which can image four centimeters at a time. The latest generation of CT technology, a 320-detector row volume CT scanner, can image 16 centimeters - or the entire length of the heart - in a single rotation and within a single heartbeat.

In his study, Dr. Einstein and a team of researchers compared the radiation exposure incurred during a coronary CT angiography procedure using a 64-detector row helical scanning and volume scanning, using a 320-detector row volume CT scanner. Phantoms simulating the male and female body were imaged using six different scan modes.

Using standard 64-detector row helical scanning as the benchmark, the effective radiation dose was reduced by 91 percent from 35.4 millisieverts (mSv) to 4.4 mSv using optimized 320-detector row volume scanning.

"By imaging the entire heart in one piece, volume scanning eliminates artifacts due to seams or gaps between image sections," said Dr. Einstein. "Moreover, the x-ray tube is left on for only a brief duration, as little as .35 seconds."

According to Dr. Einstein, state-of-the-art CT technology emphasizes optimal image resolution with the ability to lower radiation dose through a variety of features and scan modes that adjust and modulate the dose based on the specific needs of the individual patient.

"As CT technology advanced from 16- to 64-slice capabilities, the radiation dose went up significantly," he said. "Today, technology development is going in the opposite direction, reducing radiation exposure."

Dr. Einstein emphasized that practitioners must pay careful attention to using the appropriate scan mode to obtain diagnostic information with the least amount of radiation exposure to the patient.

"Radiation Dose from Single-Heartbeat Coronary CT Angiography Performed with a 320-Detector Row Volume Scanner." Collaborating with Dr. Einstein were Carl D. Elliston, M.A., Andrew E. Arai, M.D., Marcus Y. Chen, M.D., Richard Mather, Ph.D., Gregory D. N. Pearson, M.D., Ph.D., Robert L. DeLaPaz, M.D., Edward Nickoloff, D.Sc, Ajoy Dutta, M.S., and David J. Brenner, Ph.D., D.Sc.

Disclosure: Dr. Mather is an employee of Toshiba America Medical Systems. Dr. Mather provided recommendations but had no control over the data included in the manuscript.

For more information: www.rsna.org

Related Content

CT Shows Enlarged Aortas in Former Pro Football Players

3-D rendering from a cardiac CT dataset demonstrating mild dilation of the ascending aorta. Image courtesy of Christopher Maroules, M.D.

News | Computed Tomography (CT)| January 12, 2018
Former National Football League (NFL) players are more likely to have enlarged aortas, a condition that may put them at...
Siemens Healthineers Strengthens CT Portfolio With Four New Systems at RSNA 2017

The Somatom Force with the new FAST 3D Camera

Technology | Computed Tomography (CT)| December 14, 2017
December 14, 2017 — Siemens Healthineers introduced four new...
Toshiba Medical Introduces Aquilion Prime SP CT System at RSNA 2017
Technology | Computed Tomography (CT)| December 13, 2017
December 13, 2017 — Toshiba Medical, a Canon Group company, introduced its new Aquilion Prime SP...
Philips Debuts IQon Elite Spectral CT Scanner at RSNA 2017
Technology | Computed Tomography (CT)| December 11, 2017
At the 2017 Radiological Society of North America (RSNA) Annual Meeting, Philips unveiled the IQon Elite Spectral CT,...
Toshiba Highlights Ultra-High Resolution CT at RSNA 2017
News | Computed Tomography (CT)| December 06, 2017
Toshiba Medical, a Canon Group company, showcased the Aquilion Precision, what it calls the world’s first ultra-high...
Samsung Unveils Mobile CT OmniTom at RSNA 2017
Technology | Computed Tomography (CT)| November 26, 2017
Samsung Electronics debuted its OmniTom mobile 16-slice computed tomography (CT) scanner at the Radiological Society of...
Siemens Healthineers Announces First U.S. Installs of Somatom go.Up CT System
News | Computed Tomography (CT)| November 15, 2017
November 15, 2017 — Center for Diagnostic Imaging (CDI), one of the nation’s largest providers of diagnostic imaging
Medis Releases QAngio CT v3.1
Technology | Computed Tomography (CT)| October 09, 2017
Medis has released a new version of its QAngio CT (computed tomography), which can now be launched from the Medis Suite...
Hitachi Supria True64 CT Receives FDA Clearance
Technology | Computed Tomography (CT)| September 15, 2017
Hitachi Healthcare Americas Inc. announced it has attained U.S. Food and Drug Administration (FDA) 510(k) clearance to...
Orange County, Calif. Hospital Adopts Siemens Somatom Force CT for Cardiac Imaging
News | Computed Tomography (CT)| September 12, 2017
Hoag Memorial Hospital Presbyterian recently became the first hospital in Orange County, Calif., to install the Siemens...
Overlay Init