News | Atherectomy Devices | April 19, 2017

Cardiovascular Systems Initiates Voluntary Recall of Atherectomy Saline Infusion Pump

Company says electromagnetic interference may cause pumps to switch to standby made during use

CSI recall for saline pump on its atherectomy system

April 19, 2017 — Cardiovascular Systems Inc. (CSI) announced April 18 it had initiated a voluntary recall of its 7-10014 Saline Infusion Pump. CSI initiated a customer communication of the recall by letter and informed customers that they may continue to use the affected Saline Infusion Pumps until they receive a replacement.

The Saline Infusion Pumps provide saline and lubricant infusion during orbital atherectomy procedures and electrical power to the orbital atherectomy device. CSI has determined that electromagnetic interference present in the hospital environment may cause the pumps to switch to standby mode during use, requiring the pump to be reset prior to continuing treatment. Restoring pump operation can result in a temporary delay in the orbital atherectomy procedure. In coronary artery procedures, this delay of therapy could present an additional risk of a temporary, medically reversible injury. However, there have been no reports of patient injury to date.

The pumps included in the recall were distributed between April 7, 2015 and April 4, 2017. The company plans to recall and replace approximately 900 units currently in customer inventory. CSI expects to record approximately $1.5 million of expenses in its third quarter of fiscal 2017 related to the recall and replacement of all affected Saline Infusion Pumps. The recall and related charge do not have an adverse effect on third quarter revenue and net loss guidance provided by the company on January 25, 2017. The company will provide additional financial updates during its regularly scheduled third quarter earnings conference call on May 3, 2017.

For more information: www.csi360.com

Related Content

Sponsored Content | Videos | Angiography| June 22, 2017
Shimadzu Medical Systems hghlighted new features of its Trinias Interventional X-ray line at the America College of C
BioTrace Medical's Tempo Lead Shows Stability and Safety in Real-World Performance
News | Leads Implantable Devices| June 20, 2017
BioTrace Medical Inc. announced that the company’s Tempo Temporary Pacing Lead was featured in an oral presentation and...
Stroke2prevent Receives FDA 510(k) Clearance for A-View
Technology | Ultrasound Transesophageal echo (TEE)| June 20, 2017
June 20, 2017 — Dutch medical device company Stroke2prevent BV recently announced it has received U.S.
IABP recall for Maquet, Datascope intra-aortic balloon pump CS100 and CS300

The Datascope CS300 IABP console.

News | Hemodynamic Support Devices| June 20, 2017
June 20, 2017 — The U.S.
Sponsored Content | Videos | Cardiovascular Ultrasound| June 15, 2017
Madhav Swaminathan, M.D., FASC, FAHA, clinical director of cardiothoracic anesthesiology, at Duke Univer
radial access, transradial access trial using anticoagulants
News | Radial Access| June 15, 2017
June 15, 2017 — In patients undergoing t...
Mitral Valve regurgitation showed on an echocardiogram, cardiac ultrasound

Mitral valve regurgitation showed on an echocardiogram. Image from Toshiba.

News | Heart Valve Technology| June 15, 2017
June 15, 2017 — 4C Medical Technologies Inc., a developer of minimally invasive therapies for structural heart diseas
Sponsored Content | Videos | Heart Valve Technology| June 14, 2017
A discussion with Rebecca Hahn, M.D., FASE, director of interventional echocardiography and professor of medicine at
Radiation dose monitoring in cardiac imaging, cath lab angiography

An example of dose monitoring software from Sectra, which can record dose levels by exam type, modality and set alert levels. 

Feature | Radiation Dose Management| June 12, 2017 | Jeff Zagoudis
For all the benefits of medical imaging, most forms come with
Overlay Init